Agilent PVG-500S Pirani Gauge
Agilent PVG-500S Pirani Gauge Agilent / Varian Vacuum Measurement Vacuum Products Malaysia, Penang, Singapore, Indonesia Supplier, Suppliers, Supply, Supplies | Hexo Industries (M) Sdn Bhd
Active Gauges

PVG-500S Pirani Gauge

 

The Agilent PVG-500 / 500S(tungsten) / 502S(nickel) series of Pirani gauges provides the ultimate combination of state-of-art technology and ruggedness within a compact design. The PVG-500 series utilizes the most advanced digital Pirani technology combined with a stainless steel sensor design to meet the various needs of the market today.

PVG500KF16RS Replacement sensor, PVG-500 Pirani gauge, tungsten filaments, KF17
PVG500KF16 PVG-500 Pirani gauge, KF17
PVG500KF16S PVG-500 Pirani gauge, tungsten filaments, with setpoints, KF17

 
Features
  • Simplified integration with a compact, rugged aluminum housing that mounts in any orientation with a logarithmic signal output
  • Stainless steel sensor cell with metal-sealed feedthrough provides a rugged design for a wide range of applications
  • Set-up is easy with push button ATM and High Vacuum (HV) adjustment
  • Nickel filament option for applications in corrosive environments
  • Optional set points allowing for the utilization of pressure readings to perform critical operations
  •  
Specifications

Technical Specifications

Measurement range
(Air, O2, CO, N2)
5 x 10-4 to 1000 mbar (3.75 x 10-4 to 750 Torr)
Accuracy (N2) Units
% of reading + 15% 1 x 10-3 to 100 mbar (1 x 10-3 to 75 Torr)
+ 50% 5 x 10-4 to 1 x 10-3 mbar (3.75 x 10-4 to 1 x 10-3 Torr)
+ 50% 100 to 1000 mbar (75 to 750 Torr)
Repeatability (Air) 1 x 10-3 to 100 mbar (1 x 10-3 to 75 Torr) – % of reading 􏰀+2%
Output signal (measurement signal) Voltage range Measurement range
0 to 10.3 V 1.9 to 10.0 V
Voltage vs. pressure 1.286 V/decade, logarithmic
Error signal 0 to 0.5 V (filament rupture)
Minimum loaded impedance 10 kΩ, short-circuit proof
Response time 80 ms
Adjustment One tactile switch for both ATM and HV adjustment
Identification gauge 27 kΩ, referenced to supply common
Setpoint (2)   PVG-500 PVG-500S, PVG-502S
 Setting range None 2 x 10-3 to 500 mbar (1.5 x 10-3 to 375 Torr)
Hysteresis 10% above lower threshold % of reading
Relay contact 30 VDC / 0.5 ADC floating
Switching time <20 ms
Supply voltage At gauge Ripple Power consumption
14 to 30 VDC ≤1 Vpp ≤1 W
Electrical connection FCC 68/RJ45 appliance connector, 8 poles, male
Sensor cable 8 poles plus shielding
Cable length ≤ 100 meter (330 ft)
Materials exposed to vacuum Glass, Ni, NiFe, tungsten (tungsten version)
Admissible temperature Operation Storage Vacuum Connection
5 to 60 °C –20 to +65 °C 80 °C


Outline Drawing
PVG-500 Series Pirani Gauge Outline Drawing
 
Literature

Active Gauges and Controllers Brochures

This brochure contains information comparing select gauges and controllers based on your application
( https://cdn1.npcdn.net/userfiles/21330/download/1599128069_310475637543a5b9823563a6e0025b7d/Active_Gauges_and_Controllers_Brochure.pdf )



NEW PVG-500 / PVG-500S/PVG-502S Active Gauges Data Sheet