Agilent Multi-Gas CrossLab Sniffer
Agilent Multi-Gas CrossLab Sniffer Portable Sniffer Leak Detector Malaysia, Penang, Singapore, Indonesia Supplier, Suppliers, Supply, Supplies | Hexo Industries (M) Sdn Bhd G8613A, G6693A, Multi-Gas CrossLab Sniffer G6699A, Agilent GC Leak Detector G6694A, Electronic Leak Detector Cartridge
Portable Sniffer Leak Detectors

Agilent Multi-Gas CrossLab Sniffer

 

The Multi-Gas CrossLab Sniffer Leak Detector provides an industrial level of sensitivity and responsiveness for the various gas types. It is an affordable sniffing solution for detecting gas species such as helium, hydrogen, argon, and nitrogen.

This mini-sized portable leak detector is perfect for verifying and troubleshooting gas delivery systems thus avoiding costly leaks and safety concerns. It is also a perfect manufacturing tool for leak checking stations for offline leak location and repair. The battery-powered ultra lightweight ergonomic design also offers a viewable OLED display.
-

G6694-60005 Leak Detector Mesh Filter 1 Each
 
G6694-60005

Leak Detector Mesh Filter

See More Details 
G6694A Electronic Leak Detector Cartridge 1 Each
NoImage
G6694A

Electronic Leak Detector Cartridge

See More Details 
G8613A Multi-Gas CrossLab Sniffer Leak Detector 1 Each
 
G8613A

Multi-Gas CrossLab Sniffer Leak Detector

See More Details 
-
Features
  • Palm-sized, battery-powered instrument
  • Lightweight ergonomic design
  • 49.9 dB audible beep upon leak detection
  • Easy-read OLED display, 128x64 pixels monochrome w/16 grayscales
  • Bar levels displayed when desired gas is present – non quantifiable
-
Specifications
Dimensions (WxHxL)
  • 48 mm x 201 mm x 88 mm
Height
  • 201 mm
Length
  • 88 mm
Power Requirements
  • Battery (2 AA)
  • USB
Response Time
  • <2 seconds
Sensitivity
  • Helium 5E-5 std cc/sec (5.1 mbar•L/sec)
  • Hydrogen 4.2E-5 std cc/sec (4.3 mbar•L/sec)
Weight
  • 420 g
Width
  • 48 mm
Zeroing Time
  • <1.5 sec
-
How It Works

Leak-checking sniffing solution

Air is drawn in through the probe and into the manifold using a suction pump via insulated tubing. When the probe is held up to ambient air, both resistors in the sensor are cooled equally and the voltage reading can be zeroed to balance the bridge. But when the probe is held near a leak, target gases change the thermal carrying capacitance of the sample gas entering the sensor.
The intensity of the gas leak is accomplished by comparing the target gas's thermal conductivity and leak rate relative to ambient air.

Displayed leak rate

<description ng-if=''slot[attr[1]] && slot[attr[1]] != '''' ng-bind-html=''slot[attr[1]] | trustAsHtml'' ng-attr-id=''{{getId(slot,attr,1)}}'' ng-attr-insite=''{{getInsite(slot,attr,1)}}'' ''='''' id='''' insite='''' style=''box-sizing: border-box; -webkit-font-smoothing: antialiased; display: block;''>

When a gas leak is detected, bar levels are displayed on the screen and “Leak” will also appear on the screen. The number of bar levels is proportional to the size of the leak. A maximum of eight bar levels can be displayed, indicating a large volume gas leak. When the number of bar levels is three or more, the leak alert sign will blink on the screen and the buzzer will sound to give an audible alert. Two styles of bars, filled and empty, are used to indicate the type of gas entering the detector.

-
Literature

Sustainability in the Vacuum Industry: A Guide to Agilent Vacuum and Leak Detection Green Solutions

Learn more about sustainable vacuum products and solutions in this exhaustive guide.

Vacuum and Leak Detection Reference: Formulas, Properties, and Glossary

This reference helps users to keep important properties, formulas, and terms used in vacuum technology at their fingertips.

Agilent Vacuum and Leak Detection Solutions for e-Mobility

Vacuum and leak detection solutions in advanced industrial processes for vehicle manufacturing, supporting the transition to sustainable mobility.

Agilent Multi-Gas CrossLab Sniffer Leak Detector Data Sheet

Agilent Multi-Gas CrossLab Sniffer leak detector: features, technical specifications, and ordering details.

Support
Certificates of Conformance

Agilent Vacuum Product Declarations of Conformity Reference List

Contains a reference list of Vacuum Products Declarations of Conformity. Each vacuum product is shipped with the specific Declaration of Conformity.

User Manuals

Agilent CrossLab CS Electronic Leak Detector Operation Manual

Operation manual for the Agilent CrossLab CS electronic leak detector.

Agilent CrossLab CS PC Connection User Manual

This manual provides instructions on several common tasks involving Agilent CrossLab CS devices.

 
-
Contact us : sales@hexoind.com
Available in Indonesia Philippines Vietnam Singapore Thailand Malaysia Brunei Cambodia Laos Myanmar.


Please leave your enquiry here, we will reply as soon as possible.


*only support gif, jpeg, jpg, png, pdf
Switch To Desktop Version